DocumentCode
1495201
Title
Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging]
Author
Pauly, John ; Le Roux, Patrick ; Nishimura, Dwight ; Macovski, Albert
Author_Institution
Inf. Syst. Lab., Stanford Univ., CA, USA
Volume
10
Issue
1
fYear
1991
fDate
3/1/1991 12:00:00 AM
Firstpage
53
Lastpage
65
Abstract
An overview of the Shinnar-Le Roux (SLR) algorithm is presented. It is shown how the performance of SLR pulses can be very accurately specified analytically. This reveals how to design a pulse that produces a specified slice profile and allows the pulse designer to trade off analytically the parameters describing the pulse performance. Several examples are presented to illustrate the more important tradeoffs. These include linear-phase and minimum- and maximum-phase pulses. Linear-phase pulses can be refocused with a gradient reversal and can be used as spin-echo pulses. Minimum- and maximum-phase pulses have better slice profiles, but cannot be completely refocused
Keywords
biomedical NMR; patient diagnosis; NMR imaging; Shinnar-Le Roux selective excitation pulse design algorithm; gradient reversal; linear-phase pulse; maximum-phase pulse; medical diagnostic imaging; minimum-phase pulse; parameter relations; pulse performance; slice profile; Algorithm design and analysis; Biomedical imaging; Digital filters; Image resolution; Magnetic resonance imaging; Nonlinear equations; Performance analysis; Radio frequency; Slabs; Spatial resolution;
fLanguage
English
Journal_Title
Medical Imaging, IEEE Transactions on
Publisher
ieee
ISSN
0278-0062
Type
jour
DOI
10.1109/42.75611