    An Interstitial Fluid Transdermal Extraction System for Continuous Glucose Monitoring

  • Author

    Yu, Haixia ; Li, Dachao ; Roberts, Robert C. ; Xu, Kexin ; Tien, Norman C.

  • Author_Institution
    State Key Lab. of Precision Meas. Technol. & Instrum., Tianjin Univ., Tianjin, China
  • Volume
    21
  • Issue
    4
  • fYear
    2012
  • Firstpage
    917
  • Lastpage
    925
  • Abstract
    A novel microfluidic system which is fabricated with five polydimethylsiloxane layers for interstitial fluid (ISF) extraction, collection, and measurement toward the application of continuous and real-time glucose monitoring is presented in this paper. The system consists of a micro vacuum generator for ISF transdermal extraction and fluid manipulation, micro chambers for the collection of ISF, micro pneumatic valves for fluid management, and a micro flow sensor for ISF volume measurement. Sequentially controlled by the pneumatic valves, the ISF extraction, collection, and volumetric measurement functions of the system were demonstrated using the stable vacuum generated by the integrated vacuum generator. Through low-frequency ultrasound pretreated full-thickness pig skin, the normal saline solution with different glucose concentrations was transdermally extracted, collected, and measured. The absolute error in the volume measurement of the transdermally extracted “ISF analog” was less than 0.05 μL. The microfluidic system makes it possible to realize the clinical application of continuous glucose monitoring based on ISF transdermal extraction technology.
  • Keywords
    biomedical equipment; biomedical materials; biomedical ultrasonics; chemical sensors; diseases; flow sensors; microfluidics; microsensors; pneumatic actuators; polymer films; skin; sugar; ISF volume measurement; continuous glucose monitoring; diabetes; fluid management; fluid manipulation; interstitial fluid transdermal extraction system; low frequency ultrasound; microchambers; microflow sensor; microfluidic system; micropneumatic valve; microvacuum generator; polydimethylsiloxane layer; saline solution; Electrodes; Generators; Monitoring; Sugar; Vacuum systems; Valves; Volume measurement; Continuous glucose monitoring; interstitial fluid (ISF); microfluidic system; transdermal extraction;
  • fLanguage
    English
  • Journal_Title
    Microelectromechanical Systems, Journal of
  • Publisher
    ieee
  • ISSN
    1057-7157
  • Type

    jour

  • DOI
    10.1109/JMEMS.2012.2192910
  • Filename
    6189021