• DocumentCode
    1505545
  • Title

    A moving-actuator type electromechanical total artificial heart. I. Linear type and mock circulation experiments

  • Author

    Min, Byoung G. ; Kim, Hee C. ; Lee, Sang H. ; Kim, Jong W. ; Kim, Jin T. ; Kim, In Y. ; Kim, Sung W. ; Diegel, Paul D. ; Olsen, Don B.

  • Author_Institution
    Dept. of Biomed. Eng., Seoul Nat. Univ., South Korea
  • Volume
    37
  • Issue
    12
  • fYear
    1990
  • Firstpage
    1186
  • Lastpage
    1194
  • Abstract
    A novel type of motor-driven total artificial heart system with a moving-actuator mechanism has been developed. The prototype system consists of a brushless DC motor inside a rolling cylinder, two arcshaped pusher plates, and two polyurethane sacs. The moving-actuator-type electromechanical pump has the structural advantages of small size and light weight, as compared to other motor-driven pumps with fixed actuator mechanisms. Results of mock circulation tests with a cardiac output of 9 L/min at an aortic pressure of 120 mm Hg and a heart rate of 120 b/m are reported.
  • Keywords
    artificial organs; cardiology; 120 mmHg; aortic pressure; arcshaped pusher plates; brushless DC motor; cardiac output; heart rate; mock circulation experiments; motor-driven pumps; moving-actuator type electromechanical total artificial heart; polyurethane sacs; rolling cylinder; Artificial heart; Bridges; Brushless DC motors; Couplings; Heart rate; Humans; Prototypes; Testing; Vacuum systems; Wires; Algorithms; Animals; Blood Circulation; Heart, Artificial; Humans; Models, Cardiovascular; Prosthesis Design;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.64461
  • Filename
    64461