• DocumentCode
    1514895
  • Title

    Power efficient output stage for high density implantable stimulators

  • Author

    Sooksood, K. ; Ortmanns, Maurits

  • Author_Institution
    Dept. of Electron. Eng., King Mongkut´s Inst. of Technol., Bangkok, Thailand
  • Volume
    48
  • Issue
    10
  • fYear
    2012
  • Firstpage
    551
  • Lastpage
    552
  • Abstract
    In implantable stimulators, a current-controlled stimulation is preferred over a voltage-controlled one due to its safety. However, the power efficiency is a major disadvantage. By reducing the headroom voltage needed in the current driver, power efficiency of a constant current stimulation is improved. A promising technique is to bias the transistor in the triode region whereby improving output impedance through the regulated cascode structure. This comes with a feature of an implicit compliance monitor. Proposed is a new power efficient high compliance output stage.
  • Keywords
    active networks; bioelectric potentials; biomedical electronics; prosthetics; transistors; triodes; constant current stimulation; current driver; current-controlled stimulation; headroom voltage; high density implantable stimulators; implicit compliance monitor; output impedance; power efficient output stage; regulated cascode structure; transistor; triode region; voltage-controlled stimulation;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el.2012.0255
  • Filename