• DocumentCode
    1519639
  • Title

    Extension of voltage range for power and energy calibrations

  • Author

    Petersons, Oskars ; Nelson, Thomas L. ; Oldham, Nile M. ; Fitzpatrick, Gerald J.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • Volume
    48
  • Issue
    2
  • fYear
    1999
  • fDate
    4/1/1999 12:00:00 AM
  • Firstpage
    439
  • Lastpage
    443
  • Abstract
    A special purpose AC voltage divider system with ratios of 600, 480, 360, 277, and 240 V to 120 V has been developed to extend the range of primary electric power calibrations from 120-600 V at 50 and 60 Hz. The developmental goal has been to realize AC voltage scaling within 5 μV/V overall relative uncertainty. By employing a technique consisting of a special two-stage resistive divider compensated with an active circuit, the overall uncertainty achieved after calibration is 4 μV/V
  • Keywords
    active networks; bridge circuits; calibration; energy measurement; measurement standards; measurement uncertainty; operational amplifiers; power measurement; voltage dividers; 120 to 600 V; 50 Hz; 60 Hz; AC voltage divider; AC voltage scaling; active circuit compensated; energy calibration; op amp; overall uncertainty; precision measurement; primary electric power calibration; special purpose system; two-stage resistive divider; voltage range extension; Active circuits; Bridge circuits; Calibration; Circuit stability; Equations; Feedback loop; Power measurement; Resistors; Transformers; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.769620
  • Filename
    769620