• DocumentCode
    1519673
  • Title

    A method for calibration of partial discharge calibrators

  • Author

    Gunnarsson, Ove ; Bergman, Anders ; Rydler, Karl-Erik

  • Author_Institution
    Nat. Testing & Res. Inst., Boras, Sweden
  • Volume
    48
  • Issue
    2
  • fYear
    1999
  • fDate
    4/1/1999 12:00:00 AM
  • Firstpage
    453
  • Lastpage
    456
  • Abstract
    A method for calibration of partial discharge calibrators has been developed based on passive integration of the induced charge. The voltage over a known capacitor is measured by a calibrated digital sampling voltmeter and the charge is calculated as the voltage multiplied by the capacitance. The best achieved uncertainty is ±2% with a coverage factor k=2
  • Keywords
    calibration; high-voltage techniques; measurement uncertainty; partial discharge measurement; HV techniques; calibration method; charge measurement; digital sampling voltmeter; impulse charge; induced charge; partial discharge calibrators; passive integration; pulse measurements; uncertainty; Calibration; Capacitance measurement; Capacitors; Charge measurement; Current measurement; Partial discharge measurement; Partial discharges; Sampling methods; Voltage; Voltmeters;