• DocumentCode
    1528590
  • Title

    Verification of the westervelt equation for focused transducers

  • Author

    Jing, Yun ; Shen, Du ; Clement, Greg T.

  • Author_Institution
    Dept. of Radiol., Brigham and Women´´s Hosp., Boston, MA, USA
  • Volume
    58
  • Issue
    5
  • fYear
    2011
  • fDate
    5/1/2011 12:00:00 AM
  • Firstpage
    1097
  • Lastpage
    1101
  • Abstract
    This study investigates the validity of the Westervelt equation for focused transducers. The angular spectrum method is employed to analyze the second-harmonic acoustic field under the weakly nonlinear approximation. Although it is well known that the Westervelt equation is accurate for the case of quasi-plane waves, the present work demonstrates accurate solution for the highly focused case of a spherically-curved ultrasound transducer having an aperture angle of 80°. It is further found that the solution error is inversely dependent on the nonlinearity coefficient.
  • Keywords
    harmonic generation; ultrasonic transducers; Westervelt equation; angular spectrum method; focused transducer; quasi plane wave; second harmonic acoustic field; spherically curved ultrasound transducer; weakly nonlinear approximation; Acoustics; Apertures; Equations; Harmonic analysis; Mathematical model; System-on-a-chip; Transducers;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2011.1910
  • Filename
    5776764