Correction of periodic motion artifacts along the slice selection axis in MRI
Author
Mitsa, T. ; Parker, K.J. ; Smith, W.E. ; Tekalp, A.M. ; Szumowski, J.
Author_Institution
Rochester Univ., NY, USA
Volume
9
Issue
3
fYear
1990
fDate
9/1/1990 12:00:00 AM
Firstpage
310
Lastpage
317
Abstract
The effect of periodic motion of a single magnetic resonance imaging (MRI) slice in the direction of the slice selection axis is modeled as amplitude modulation of the raw data with a motion kernel along the phase encoding direction in the Fourier domain. It is shown that this motion can be detected in 1-D projections of the raw data along the frequency encoding direction which in combination with appropriate filtering leads to the recovery of the motion kernel. It is demonstrated by means of simulation examples that significant reduction in the amplitude of ghost artifacts is obtained when the image is filtered by the inverse of the motion kernel. Some issues to be investigated before the technique can be used in a clinical environment are mentioned
Keywords
biomedical NMR; patient diagnosis; Fourier domain; NMR imaging; artifact correction; clinical environment; ghost artifacts; image filtering; magnetic resonance imaging; medical diagnostic imaging; motion kernel; periodic motion artifacts; phase encoding direction; raw data amplitude modulation; slice selection axis; Amplitude modulation; Data acquisition; Encoding; Filtering; Frequency; Kernel; Magnetic resonance imaging; Motion detection; Optical filters; Optical imaging; Optical sensors; Radiology;
fLanguage
English
Journal_Title
Medical Imaging, IEEE Transactions on
Publisher
ieee
ISSN
0278-0062
Type
jour
DOI
10.1109/42.57769
Filename
57769
Link To Document