DocumentCode
1587274
Title
Pivot shift evaluation in the knee joint using inertial and magnetic sensors
Author
Tsumori, Y. ; Yoshiya, Shinich ; Yagi, Masashi ; Shibanuma, Nao ; Tomaru, A. ; Kobashi, Shoji ; Kondo, K. ; Hata, Yuki
Author_Institution
Dept. of Orthopaedic Surg., Hyogo Coll. of Med., Nishinomiya, Japan
fYear
2010
Firstpage
1
Lastpage
4
Abstract
The pivot shift test has been performed to assess the instability of the knee caused by anterior cruciate ligament (ACL) injuries. However, the test depends on the clinician´s subjective feeling. In this study, inertial and magnetic sensors have been introduced into quantitative evaluation of the pivot shift test. The analysis method extracts the knee movement of the pivot shift by using wavelet transformation. In the result of applying the proposed method to the ACL injured subject, pivot shift phenomenon was detected correctly in comparison with reference video images taken simultaneously, the mean of the maximum accelerations of the pivot shift was 2.19 ± 0.69 m/s2, and the maximum accelerations were correlated with grade scores based on the clinician´s subjective feeling.
Keywords
biomedical equipment; bone; injuries; magnetic sensors; patient diagnosis; wavelet transforms; anterior cruciate ligament injuries; clinician subjective feeling; inertial sensor; knee joint; knee movement; magnetic sensor; maximum accelerations; pivot shift test; reference video images; wavelet transformation; Anterior Cruciate Ligament; Inertial and magnetic sensor; Kinematics; Knee joint; Pivot shift; Quantification; Wavelet;
fLanguage
English
Publisher
ieee
Conference_Titel
World Automation Congress (WAC), 2010
Conference_Location
Kobe
ISSN
2154-4824
Print_ISBN
978-1-4244-9673-0
Electronic_ISBN
2154-4824
Type
conf
Filename
5665337
Link To Document