ADONIS, high count-rate HP-Ge γ spectrometry algorithm: Irradiated fuel assembly measurement
Author
Pin, Patrick ; Barat, Eric ; Dautremer, Thomas ; Montagu, Thierry ; Normand, Stéphane
Author_Institution
Nucl. Meas. Team, AREVA NC La Hague, Beaumont-Hague, France
fYear
2011
Firstpage
1
Lastpage
5
Abstract
ADONIS is a digital system for gamma-ray spectrometry, developed by CEA. This system achieves high count-rate gamma-ray spectrometry with correct dynamic dead-time correction, up to, at least, more than an incoming count rate of 3.106 events per second. An application of such a system at AREVA NC´s La Hague plant is the irradiated fuel scanning facility before reprocessing. The ADONIS system is presented, then the measurement set-up and, last, the measurement results with reference measurements.
Keywords
gamma-ray spectrometers; germanium radiation detectors; ADONIS digital system; HP-Ge gamma spectrometry algorithm; dynamic dead-time correction; high count-rate gamma-ray spectrometry; high-purity germanium detector; irradiated fuel assembly measurement; irradiated fuel scanning facility; Apertures; Assembly; Detectors; Fuels; Instruments; Signal processing algorithms; Spectroscopy; gamma-ray spectrometry; high count-rate; high-purity germanium detector (HP-Ge); irradiated fuel measurement;