• DocumentCode
    1614156
  • Title

    Validation of electron beam position monitoring probes by comparison of dose profile measurements with beam displacement sensing in the microwave range

  • Author

    Ruf, Marcel ; Schmidt, Lorenz-Peter ; Setzer, Stefan

  • Author_Institution
    High Freq. Technol., Univ. of Erlangen-Nuremberg, Nuremberg, Germany
  • fYear
    2011
  • Firstpage
    1410
  • Lastpage
    1413
  • Abstract
    The development of a beam position monitoring and control architecture for compact linear accelerators requires novel sensors which fulfill the tight space restrictions given in such systems. This paper presents simulation and experimental data of capacitive pickup probes with high sensitivity capable of sensing beam deflections far below 1 mm and compact enough to be integrated in accelerating structures. The correlation of beam position monitoring output signals with dose profile measurements is presented indicating that the proposed compact sensors together with dedicated BPM hardware allow reliable electron beam trajectory determination.
  • Keywords
    dosimetry; electron beams; linear accelerators; microwave spectra; position measurement; trajectory control; BPM hardware; beam displacement sensing; capacitive pickup probes; compact linear accelerators; control architecture; correlation; dose profile measurements; electron beam position monitoring probes; electron beam trajectory; microwave range; sensing beam deflections; Current measurement; Linear accelerators; Monitoring; Particle beams; Position measurement; Probes; Sensitivity; Beam Position Monitoring; Electron Linacs;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference Proceedings (APMC), 2011 Asia-Pacific
  • Conference_Location
    Melbourne, VIC
  • Print_ISBN
    978-1-4577-2034-5
  • Type

    conf

  • Filename
    6174025