A technique for measuring the energy response of a PET tomograph using a compact scattering source
Author
Watson, C.C.
Author_Institution
CTI PET Syst. Inc., Knoxville, TN, USA
Volume
3
fYear
1996
Firstpage
1562
Abstract
A technique is presented for verifying the energy calibration of PET tomographs by measuring the energy response spectrum of its detectors. Single scattered radiation provides a continuous energy reference source of coincident photons. The response of the scanner vs. scattering angle, and hence vs. energy is recorded within the sinogram. From these data, the position and resolution of discriminator levels which are set below 511 keV can be extracted. This is shown to be a useful diagnostic of the scanner´s energy calibration, and of its response to scatter in 3D acquisitions
Keywords
biomedical equipment; biomedical measurement; calibration; gamma-ray scattering; positron emission tomography; 3D acquisitions; 511 keV; PET tomograph energy response measurement technique; PET tomographs energy calibration verification; compact scattering source; discriminator levels; medical diagnostic imaging; medical instrumentation; nuclear medicine; scattering angle; sinogram; Calibration; Detectors; Distributed computing; Electromagnetic scattering; Energy measurement; Energy resolution; Event detection; Particle scattering; Positron emission tomography; Pulse measurements;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
Conference_Location
Anaheim, CA
ISSN
1082-3654
Print_ISBN
0-7803-3534-1
Type
conf
DOI
10.1109/NSSMIC.1996.587923
Filename
587923
Link To Document