• DocumentCode
    1646806
  • Title

    Pressure detection and wireless interface for patient bed

  • Author

    Manohar, Achutan ; Bhatia, Dinesh

  • Author_Institution
    Electr. Eng. Dept, Univ. of Texas at Dallas, Richardson, TX
  • fYear
    2008
  • Firstpage
    389
  • Lastpage
    392
  • Abstract
    A low power, microcontroller based patient bed monitoring system is discussed in this paper. The resistive bend sensor based approach provides an effective low cost alternative for minimizing the harmful effects of bedsores which is prevalent in hospital intensive care units and assisted living environments during rehabilitation. The ZigBee based wireless interface for this pressure detection system is aimed at reducing the hassles of the existing complex wired setup and to provide continuous patient care and monitoring by enabling the patient bed mobility as essentially required for patients in critical condition.
  • Keywords
    biomedical electronics; biomedical equipment; health care; microcontrollers; patient monitoring; patient rehabilitation; pressure sensors; wireless sensor networks; wounds; ZigBee method; bedsore; critical condition monitoring; intensive care unit; low power patient bed monitoring system; microcontroller-based system; pressure detection system; resistive bend sensor; wireless interface system; Back; Blades; Bones; Costs; Hospitals; Microcontrollers; Patient monitoring; Wheelchairs; Wireless sensor networks; ZigBee; Bedsore; Decubitus ulcer; Patient bed; Pressure detection; Pressure sore; Pressure ulcer; Wireless Sensor Networks; ZigBee;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference, 2008. BioCAS 2008. IEEE
  • Conference_Location
    Baltimore, MD
  • Print_ISBN
    978-1-4244-2878-6
  • Electronic_ISBN
    978-1-4244-2879-3
  • Type

    conf

  • DOI
    10.1109/BIOCAS.2008.4696956
  • Filename
    4696956