DocumentCode
1664386
Title
Optimization of a Biomedical Actuator for Implantable Continuous Glucose Monitoring
Author
Merzaghi, Sebastiano ; Stefanini, Igor ; Markovic, Miroslav ; Perriard, Yves
Author_Institution
Ecole Polytech. Fed. de Lausanne, Lausanne
fYear
2007
Firstpage
869
Lastpage
874
Abstract
An optimization of a biomedical actuator for implantable continuous glucose concentration monitoring is presented. The actuator has a stator with a 3-phase winding and a cylindrical permanent magnet as the rotor, which is plunged in the fluid in which the glucose concentration is to be measured. The fluid viscosity, and therefore the resistive torque on the rotor, is proportional to the glucose concentration. This torque is computed by measuring the back EMF induced in the open stator phases during the deceleration of the rotor. The optimization is based on the actuator analytical model, combined with an advanced SQP algorithm.
Keywords
microactuators; permanent magnet machines; prosthetics; rotors; stators; torque; viscosity; biomedical actuators; fluid viscosity; implantable countinuous glucose concentration monitoring; permanent magnet machines; resistive torque; Actuators; Analytical models; Biomedical measurements; Biomedical monitoring; Permanent magnets; Phase measurement; Stator windings; Sugar; Torque measurement; Viscosity;
fLanguage
English
Publisher
ieee
Conference_Titel
Industry Applications Conference, 2007. 42nd IAS Annual Meeting. Conference Record of the 2007 IEEE
Conference_Location
New Orleans, LA
ISSN
0197-2618
Print_ISBN
978-1-4244-1259-4
Electronic_ISBN
0197-2618
Type
conf
DOI
10.1109/07IAS.2007.136
Filename
4347885
Link To Document