• DocumentCode
    1700367
  • Title

    Depth measurement in a strip detector

  • Author

    Wulf, E.A. ; Ampe, J. ; Johnson, W.N. ; Kroeger, R.A. ; Kurfess, J.D. ; Phlips, B.F.

  • Author_Institution
    Nat. Res. Council, Naval Res. Lab., Washington, DC, USA
  • Volume
    1
  • fYear
    2001
  • Firstpage
    58
  • Abstract
    We have demonstrated the ability to determine the depth of the gamma-ray interaction point in a thick germanium strip detector. This capability provides depth resolution to less than 0.5 mm in a device with 2 mm strip pitch. We have developed electronics that implement three dimensional readout of an entire 25×25 germanium strip detector. These electronics consume 425 mW/channel and can be easily replicated for multiple detectors. Experiments examining this systems capabilities and demonstrating a simple Compton telescope using one detector have been performed.
  • Keywords
    germanium radiation detectors; position sensitive particle detectors; 2.0 mm; Compton telescope; Ge; Ge strip detector; depth measurement; depth resolution; electronics; gamma-ray interaction point; Charge carrier processes; Energy resolution; Face detection; Gamma ray detection; Gamma ray detectors; Germanium; Laboratories; Strips; Telephony; Timing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2001 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-7324-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2001.1008409
  • Filename
    1008409