• DocumentCode
    1707876
  • Title

    Automatic monitoring of the infusion system in a rotary heart assist device

  • Author

    Yu, Yih-Choung

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Lafayette Coll., Easton, PA, USA
  • fYear
    2003
  • Firstpage
    116
  • Lastpage
    117
  • Abstract
    A fault detection algorithm was designed to monitor the infusion system of a ventricular assist device (AB-180 Circulatory Support System [AB-180 CSS], CardiacAssist, Inc.). This algorithm measures the infusion pressure, calculates the pressure derivative, and then compares to the preset thresholds to detect leaks and kinks in the infusion line, abnormal infusion rate, and infusion pump stoppage. The thresholds were identified from an experiment simulating the possible fault conditions of the infusion system. The algorithm was implemented in a data acquisition system and verified by using different pumps and controllers.
  • Keywords
    artificial organs; blood pressure measurement; cardiology; computerised monitoring; patient monitoring; pumps; AB-180 Circulatory Support System; abnormal infusion rate; algorithm; controllers; data acquisition system; fault conditions; fault detection algorithm; infusion line; infusion pressure; infusion pump stoppage; infusion system automatic monitoring; kink detection; leak detection; preset thresholds; pressure derivative; pumps; rotary blood pumps; rotary heart assist device; thresholds; ventricular assist device; Algorithm design and analysis; Cascading style sheets; Computerized monitoring; Control systems; Data acquisition; Fault detection; Fault diagnosis; Heart; Leak detection; Pressure measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioengineering Conference, 2003 IEEE 29th Annual, Proceedings of
  • Print_ISBN
    0-7803-7767-2
  • Type

    conf

  • DOI
    10.1109/NEBC.2003.1216019
  • Filename
    1216019