DocumentCode
1707876
Title
Automatic monitoring of the infusion system in a rotary heart assist device
Author
Yu, Yih-Choung
Author_Institution
Dept. of Electr. & Comput. Eng., Lafayette Coll., Easton, PA, USA
fYear
2003
Firstpage
116
Lastpage
117
Abstract
A fault detection algorithm was designed to monitor the infusion system of a ventricular assist device (AB-180 Circulatory Support System [AB-180 CSS], CardiacAssist, Inc.). This algorithm measures the infusion pressure, calculates the pressure derivative, and then compares to the preset thresholds to detect leaks and kinks in the infusion line, abnormal infusion rate, and infusion pump stoppage. The thresholds were identified from an experiment simulating the possible fault conditions of the infusion system. The algorithm was implemented in a data acquisition system and verified by using different pumps and controllers.
Keywords
artificial organs; blood pressure measurement; cardiology; computerised monitoring; patient monitoring; pumps; AB-180 Circulatory Support System; abnormal infusion rate; algorithm; controllers; data acquisition system; fault conditions; fault detection algorithm; infusion line; infusion pressure; infusion pump stoppage; infusion system automatic monitoring; kink detection; leak detection; preset thresholds; pressure derivative; pumps; rotary blood pumps; rotary heart assist device; thresholds; ventricular assist device; Algorithm design and analysis; Cascading style sheets; Computerized monitoring; Control systems; Data acquisition; Fault detection; Fault diagnosis; Heart; Leak detection; Pressure measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioengineering Conference, 2003 IEEE 29th Annual, Proceedings of
Print_ISBN
0-7803-7767-2
Type
conf
DOI
10.1109/NEBC.2003.1216019
Filename
1216019
Link To Document