• DocumentCode
    171470
  • Title

    Time dependent electrical skin model: For the testing of electrocutaneous stimulating electrodes

  • Author

    Medeiros, Courtney ; DeSanto, Christopher Ross ; McDonough, Ryan ; Chabot, Eugene ; Ying Sun

  • Author_Institution
    Dept. of Electr., Comput., & Biomed. Eng., Univ. of Rhode Island, Kingston, RI, USA
  • fYear
    2014
  • fDate
    25-27 April 2014
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    Skin resistance changes dynamically during the time course of an external electrical stimulation. An embedded system was developed to provide a dynamic impedance in response to external stimulations. A microprocessor adjusts the series resistance component of an impedance based on a preprogrammed function. This skin impedance model is useful for testing devices involving electrocutaneous stimulation or functional electrical stimulation.
  • Keywords
    bioelectric potentials; biomedical electronics; biomedical equipment; electric resistance; embedded systems; skin; dynamic impedance; electrocutaneous stimulation; embedded system; external electrical stimulation; external stimulations; functional electrical stimulation; microprocessor; preprogrammed function; series resistance component; skin impedance model; skin resistance; time dependent electrical skin model; Electrodes; Immune system; Impedance; Microprocessors; Resistance; Skin; Table lookup; electrocutaneous stimulation; skin impedance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioengineering Conference (NEBEC), 2014 40th Annual Northeast
  • Conference_Location
    Boston, MA
  • Type

    conf

  • DOI
    10.1109/NEBEC.2014.6972873
  • Filename
    6972873