• DocumentCode
    1718118
  • Title

    Feedback control of high-intensity silicon ultrasonic surgical actuator

  • Author

    Son, Il-Seok ; Lal, Amit

  • Author_Institution
    SonicMEMS Lab., Wisconsin Univ., Madison, WI, USA
  • Volume
    1
  • fYear
    2005
  • Firstpage
    252
  • Abstract
    For the first time, we demonstrate feedback control of a high-intensity silicon ultrasonic horn actuator in the presence of time-varying contact forces, by using strain feedback from integrated piezoresistive strain gauges. With feedback control, the ultrasonic actuator has maintained its resonance under forces as high as 1.2 N on the tip of the actuator. This robust actuation enables the silicon ultrasonic horn to maintain its most effective actuation when its environment and loading conditions are changing due to ultrasonic effects such as pumping, cutting, and atomizing.
  • Keywords
    actuators; feedback; piezoresistive devices; silicon; strain gauges; strain sensors; ultrasonic transducers; feedback control; high-intensity silicon ultrasonic horn actuator; integrated piezoresistive strain gauges; strain feedback; time-varying contact forces; ultrasonic atomizing; ultrasonic cutting; ultrasonic pumping; Actuators; Capacitive sensors; Feedback control; Piezoresistance; Resonance; Resonant frequency; Silicon; Strain measurement; Surgery; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems, 2005. Digest of Technical Papers. TRANSDUCERS '05. The 13th International Conference on
  • Print_ISBN
    0-7803-8994-8
  • Type

    conf

  • DOI
    10.1109/SENSOR.2005.1496404
  • Filename