    Miniaturised annular transducers built using ultrasound micromachining for medical imaging application

  • Author

    Ballandras, S. ; Guine, P. ; Maître, P. ; Steichen, W. ; Piranda, B. ; Gelly, J.F.

  • Author_Institution
    Lab. de Phys. et Metrol. des Oscillateurs, CNRS, Besancon, France
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    192
  • Lastpage
    195
  • Abstract
    Fabrication and test of a 2 mm diameter 4 channel annular array is described. Comparison between experimentally measured and theoretically predicted admittance of the transducer elements is reported. The theoretical 3D finite element analysis is finally used to identify the electrically measured vibration mode of the device
  • Keywords
    biomedical transducers; biomedical ultrasonics; finite element analysis; micromachining; ultrasonic transducer arrays; 2 mm; electrically measured vibration mode; experimentally measured admittance; medical imaging application; medical instrumentation; miniaturised annular transducers; theoretical 3D finite element analysis; theoretically predicted admittance; transducer elements; ultrasound micromachining; Biomedical imaging; Biomedical transducers; Fabrication; Finite element methods; Micromachining; Piezoelectric transducers; Testing; Ultrasonic imaging; Ultrasonic transducers; Ultrasonic variables measurement;