• DocumentCode
    1741124
  • Title

    The fabrication of conformal ultrasound phased arrays for thermal therapy

  • Author

    McGough, Robert J. ; Owens, Amanda M. ; Cindric, Davor ; Heim, Jon W. ; Samulski, Thaddeus V.

  • Author_Institution
    Dept. of Radiat. Oncology, Duke Univ. Med. Center, Durham, NC, USA
  • Volume
    3
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    1617
  • Abstract
    A method for fabricating two-dimensional (2D) ultrasound phased arrays for thermal therapy is described. A 96 element conformal prototype is constructed, and focusing results are also presented. Measured focal patterns indicate that these conformal arrays require some calibration before focusing. Results show that offsets in the element coordinates with respect to the expected positions are the primary source of focal degradation. Measured focal patterns improve dramatically after compensating for element offsets and variations in electronic component values
  • Keywords
    biomedical transducers; biomedical ultrasonics; calibration; hyperthermia; radiation therapy; ultrasonic focusing; ultrasonic transducer arrays; 2D ultrasound phased arrays; 96 element conformal prototype; calibration; conformal ultrasound phased arrays; electronic component values; element coordinate offsets; element offsets; fabrication; focal degradation; focal patterns; focusing results; thermal therapy; Applicators; Calibration; Fabrication; Heating; Hyperthermia; Medical treatment; Phased arrays; Prototypes; Ultrasonic imaging; Ultrasonic variables measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-6465-1
  • Type

    conf

  • DOI
    10.1109/IEMBS.2000.900384
  • Filename