• DocumentCode
    1741146
  • Title

    Monte Carlo simulation of the effective point of measurement of the Markus chamber for electrons, the upright PinPoint chamber for photons, and other chambers

  • Author

    Rosenow, Ulf F. ; Thieke, Christian ; Lauterbach, Marc

  • Author_Institution
    Clinical Radiat. Phys., Georg-August-Univ., Gottingen, Germany
  • Volume
    3
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    1697
  • Abstract
    Monte Carlo simulation is shown to provide an ideal tool for the study of effective point of measurement (Peff) shifts in some electron chambers such as the Markus chamber at all energies or chambers of 2 mm plate separation at very low electron energies. Furthermore it enables the definition of Peff for the PinPoint chamber in the upright orientation in photon and electron beams
  • Keywords
    Monte Carlo methods; dosimeters; dosimetry; electron detection; gamma-ray detection; ionisation chambers; Markus chamber; Monte Carlo simulation; NACP chamber; Roos chamber; depth-dose curves; dosimetry; effective point of measurement; electron beams; electron chambers; ionisation chambers; photon beams; photon chambers; upright PinPoint chamber; Cancer; Computational biology; Computational modeling; Electromagnetic scattering; Electron beams; Imaging phantoms; Ionization chambers; Linux; Particle scattering; Physics;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-6465-1
  • Type

    conf

  • DOI
    10.1109/IEMBS.2000.900407
  • Filename