Safety Investigation of a Magnetic Pulse Applicator for Heart Stimulation
Author
De Leo, A. ; Primiani, V. Mariani ; Russo, P. ; Moglie, Franco ; Cerri, G.
Author_Institution
Dipt. di Ing. dell´Inf., Univ. Politec. delle Marche, Ancona, Italy
Volume
50
Issue
7
fYear
2014
fDate
41821
Firstpage
1
Lastpage
8
Abstract
The safety area around the applicator of a high current pulse generator, able to magnetically stimulate the human heart, is characterized in terms of allowable induced current. The magnetic field produced by the adopted coil is computed using a complete analytical model to determine the limit area allowed for medical operators. The International Commission on Non-Ionizing Radiation Protection recommendation is adopted for the occupational exposure limits. The presence of a ferrite tile close to the stimulation coil is also investigated. The ferrite tile enhances the field level on the patient side, making the magnetic stimulation more efficient. Experimental validation demonstrates the model´s reliability.
Keywords
biomedical equipment; cardiology; ferrite devices; magnetic fields; safety; International Commission on Non-Ionizing Radiation Protection; ferrite tile; heart stimulation; magnetic field; magnetic pulse applicator; occupational exposure limits; stimulation coil; Biomedical equipment; Ferrites; Heart; Magnetic domains; Magnetic field measurement; Magnetic fields; Human exposure; magnetic stimulation; pulsed magnetic field;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/TMAG.2014.2305135
Filename
6734695
Link To Document