DocumentCode
1779061
Title
Measuring Harmonic Distortion Accurately Using Spectrum Analyzer
Author
Cao Yun ; Qiu Xinyu ; Zhang Xiyang
Author_Institution
Dept. of Meas., Beijing Aerosp. Control Center, Beijing, China
fYear
2014
fDate
18-20 Sept. 2014
Firstpage
920
Lastpage
923
Abstract
The measurement result might incorrect if parameter setting is unsuitable when using spectrum analyzer to measure a signal´s harmonic distortion. It is especially useful to make sure the dynamic range of the spectrum analyzer. According to the definite dynamic range, the mixer input level and attenuator setting can be known. By analyzing the example of harmonic distortion measurement using spectrum analyzer, the accurate measurement method is given.
Keywords
distortion measurement; harmonic distortion; measurement systems; spectral analysers; attenuator setting; dynamic range; harmonic distortion measurement; mixer input level; spectrum analyzer; Attenuation measurement; Attenuators; Distortion measurement; Dynamic range; Harmonic distortion; Measurement uncertainty; Mixers; attenuator setting; dynamic range; harmonic distortion; mixer level; spectrum analyzer;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement, Computer, Communication and Control (IMCCC), 2014 Fourth International Conference on
Conference_Location
Harbin
Print_ISBN
978-1-4799-6574-8
Type
conf
DOI
10.1109/IMCCC.2014.193
Filename
6995163