    Observations of the choice of reconstruction matrix in magnetic resonance imaging

  • Author

    Young, I.R. ; Hall, A.S

  • Author_Institution
    GEC Res. Ltd., Wembley, UK
  • fYear
    1988
  • fDate
    14-17 Nov 1988
  • Firstpage
    1187
  • Abstract
    A number of ways are mentioned in which the choice of imaging matrix can affect the performance of a machine in magnetic resonance imaging. It is observed that, contrary to most other techniques, there are sensitivity and detectability advantages, in some instances, in lower-resolution images which may not be obtainable in higher-resolution ones-although it is equally true that there are other circumstances where the reverse is true. Considerations of effects arising from field inhomogeneity, patient motion, and fluid flow are used to illustrate the argument
  • Keywords
    biomedical NMR; computerised picture processing; magnetic resonance spectroscopy; computerised picture processing; detectability; field inhomogeneity; fluid flow; magnetic resonance imaging; medical NMR; patient motion; reconstruction matrix; sensitivity; Data acquisition; Fluid flow; High-resolution imaging; Image reconstruction; Image resolution; Magnetic resonance imaging; Magnets; Radiology; Signal processing; Spatial resolution;