Abstract
Several methods for deriving spatial information are tested for block configuration detectors. The performances of detectors based on continuous scintillator blocks and a module based on pixellated scintillators are compared. We showed that undivided scintillator block designs enables better spatial resolution performance than the pixellated detector. In addition, as attested by our results for incidence angles up to 40° with respect to the normal direction, contrary to the pixellated detector, the block configuration allows correcting for depth-of-interaction (DOI).
Keywords
avalanche photodiodes; gamma-ray detection; image resolution; medical image processing; positron emission tomography; solid scintillation detectors; APD array; PET detector; block configuration detectors; continuous LSO scintillator; pixellated scintillators; spatial resolution; Animal structures; Costs; Crystalline materials; Detectors; Photodetectors; Positron emission tomography; Sensor arrays; Space technology; Spatial resolution; Testing;