• DocumentCode
    1826992
  • Title

    Performance evaluation of a PET detector module using a continuous LSO scintillator coupled to an APD array

  • Author

    Léonard, S. ; Bruyndonckx, P. ; Tavernier, S.

  • Author_Institution
    Vrije Univ., Brussels, Belgium
  • Volume
    3
  • fYear
    2003
  • fDate
    19-25 Oct. 2003
  • Firstpage
    2160
  • Abstract
    Several methods for deriving spatial information are tested for block configuration detectors. The performances of detectors based on continuous scintillator blocks and a module based on pixellated scintillators are compared. We showed that undivided scintillator block designs enables better spatial resolution performance than the pixellated detector. In addition, as attested by our results for incidence angles up to 40° with respect to the normal direction, contrary to the pixellated detector, the block configuration allows correcting for depth-of-interaction (DOI).
  • Keywords
    avalanche photodiodes; gamma-ray detection; image resolution; medical image processing; positron emission tomography; solid scintillation detectors; APD array; PET detector; block configuration detectors; continuous LSO scintillator; pixellated scintillators; spatial resolution; Animal structures; Costs; Crystalline materials; Detectors; Photodetectors; Positron emission tomography; Sensor arrays; Space technology; Spatial resolution; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352308
  • Filename
    1352308