Development of real-time motion artifact reduction algorithm for a wearable photoplethysmography
Author
Hyonyoung Han ; Min-Joon Kim ; Jung Kim
Author_Institution
Korea Adv. Inst. of Sci. & Technol., Daejeon
fYear
2007
fDate
22-26 Aug. 2007
Firstpage
1538
Lastpage
1541
Abstract
This paper presents a motion artifact reduction algorithm for a real-time, wireless and wearable photoplethysmography (PPG) device for measuring heart beats. A wearable finger band PPG device consists of a 3-axis accelerometer, infrared LED, photo diode, a microprocessor and wireless module. Sources of the motion artifacts were investigated from the hand motions, through computing the correlations between the three directional finger motions and distorted PPG signals. A two-dimensional active noise cancellation algorithm was applied to compensate the distorted signals by motions, using the directional accelerometer data. NLMS (Normalized Least Mean Square) adaptive filter (4th order) was employed in the algorithm. As a result, the signals´ distortion rates were reduced from 52.34% to 3.53%, at frequencies between 1 and 2.5 Hz, which representing daily motions such walking and jogging. The wearable health monitoring device equipped with the motion artifact reduction algorithm can be integrated as a terminal in a so-called ubiquitous healthcare system, which provides a continuous health monitoring without interrupting a daily life.
Keywords
active filters; biomedical equipment; distortion; gait analysis; health care; interference suppression; least mean squares methods; medical signal processing; patient monitoring; plethysmography; NLMS; accelerometer; active noise cancellation algorithm; adaptive filter; continuous health monitoring; frequency 1 Hz to 2.5 Hz; hand motion; heart beat measurement; infrared LED; microprocessor; normalized least mean square; photo diode; real-time motion artifact reduction algorithm; signal distortion; ubiquitous healthcare system; wearable finger band PPG device; wearable health monitoring device; wearable photoplethysmography; wireless PPG device; Accelerometers; Adaptive filters; Biomedical monitoring; Distortion measurement; Fingers; Heart beat; Light emitting diodes; Microprocessors; Motion measurement; Noise cancellation; Algorithms; Artifacts; Diagnosis, Computer-Assisted; Heart Rate; Humans; Monitoring, Ambulatory; Motion; Photoplethysmography; Reproducibility of Results; Sensitivity and Specificity;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE
Conference_Location
Lyon
ISSN
1557-170X
Print_ISBN
978-1-4244-0787-3
Type
conf
DOI
10.1109/IEMBS.2007.4352596
Filename
4352596
Link To Document