• DocumentCode
    1828469
  • Title

    Equivalent perception dose of pulsed fluoroscopy

  • Author

    Aufrichtig, Richard ; Xue, Ping ; Wilson, David L.

  • Author_Institution
    Dept. of Biomed. Eng. & Radiol., Case Western Reserve Univ., Cleveland, OH, USA
  • fYear
    1994
  • fDate
    3-6 Nov 1994
  • Firstpage
    526
  • Abstract
    The authors find that pulsed fluoroscopy of stationary objects gives average dose savings of 22%, 38%, and 49% at 15, 10 and 7.5 acq/sec, respectively. Dose savings depend on object size, with less savings for smaller objects. A model that includes the spatio-temporal response of the human visual system describes these results with one free parameter
  • Keywords
    diagnostic radiography; dosimetry; visual perception; equivalent perception dose; human visual system; medical diagnostic imaging; object size; pulsed fluoroscopy; spatiotemporal response; stationary objects; Filtering; Frequency; Humans; Image sequences; Matched filters; Object detection; Radiology; Signal to noise ratio; Visual system; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE
  • Conference_Location
    Baltimore, MD
  • Print_ISBN
    0-7803-2050-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1994.411921
  • Filename
    411921