Estimating an acquisition-specific NEC curve for PET acquisitions
Author
Stearns, Charles W.
Author_Institution
GE Med. Syst., Waukesha, WI, USA
Volume
4
fYear
2003
fDate
19-25 Oct. 2003
Firstpage
2578
Abstract
NEC (or NECR) analysis is a valuable tool in understanding PET image quality and imaging performance. NEC is often expressed as a function of activity level, measured in the course of a decay series acquisition. In patient imaging a series of acquisitions is generally unavailable, and the NEC curve cannot be appreciated. This work uses the prompt and delayed coincidence event rates measured in a PET data frame, plus the measurements of system deadtime made in the course of that acquisition, to determine the shape of the NEC curve. This will allow, for instance, analysis of the image quality effects of changing the activity concentration in the patient, either by changing the injected dose or the delay to scan start.
Keywords
dosimetry; medical image processing; noise; positron emission tomography; NECR analysis; PET acquisitions; PET image quality; acquisition-specific NEC curve; activity concentration; delayed coincidence event rates; image quality effects; imaging performance; injected dose; patient imaging; Delay; Detectors; Image analysis; Image quality; Imaging phantoms; National electric code; Performance analysis; Positron emission tomography; Scattering; Shape measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2003 IEEE
ISSN
1082-3654
Print_ISBN
0-7803-8257-9
Type
conf
DOI
10.1109/NSSMIC.2003.1352417
Filename
1352417
Link To Document