Error Correction and Power Calibration of Active Load Pull Measurement Systems
Author
Teeter, Douglas A. ; East, Jack R. ; Haddad, George I.
Author_Institution
Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI 48109-2122
Volume
19
fYear
1991
fDate
33390
Firstpage
86
Lastpage
93
Abstract
A 26.5 to 40 GHz active load pull measurement system has been developed using a commercially available test fixture and bias tees. At 27 GHz, the system is capable of reflection coefficient magnitude accuracy better than 0.007, phase error less than 4 degrees, and gain uncertainty less than 0.15 dB. Reflection coefficient measurements are corrected by flrst using a standard network analyzer TRL calibration routine and then converting it to two one port calibrations. Correction of the power measurements is not as straight forward. Sources of error in the power measurements and methods to reduce this error are discussed.
Keywords
Calibration; Electric variables measurement; Equations; Error correction; Frequency; Lifting equipment; Power generation; Power measurement; Reflection; System testing;