DocumentCode
1840247
Title
A Convenient Indication of the Effect of Instrumentation Errors on Six - Port Reflection Measurements
Author
Potter, Chris
Author_Institution
Marconi Instruments, Stevenage, England
Volume
26
fYear
1994
fDate
1-2 Dec. 1994
Firstpage
10
Lastpage
19
Abstract
The performance of a six - port Vector Reflection Analyzer is mainly limited by the accuracy of its calibration kit, but the fundamental limit is due to the instrumentation. This paper shows the relationship between instrumentation errors in scalar power measurements and the corresponding residual directivity, source match and frequency response errors in a six - port reflection measurement. Furthermore, enough measurements are made during a normal calibration to produce a figure of merit for the instrumentation as a by-product.
Keywords
Calibration; Circuits; Computer errors; Detectors; Frequency response; Instruments; Performance analysis; Power measurement; Radio frequency; Reflection;
fLanguage
English
Publisher
ieee
Conference_Titel
ARFTG Conference Digest-Fall, 44th
Conference_Location
Boulder, CO, USA
Print_ISBN
0-7803-5686-1
Type
conf
DOI
10.1109/ARFTG.1994.327076
Filename
4119751
Link To Document