• DocumentCode
    1841955
  • Title

    Wearable Seismocardiography

  • Author

    Castiglioni, P. ; Faini, A. ; Parati, G. ; Di Rienzo, M.

  • Author_Institution
    Polo Tecnol., Milan
  • fYear
    2007
  • fDate
    22-26 Aug. 2007
  • Firstpage
    3954
  • Lastpage
    3957
  • Abstract
    Seismocardiogram (SCG) is the recording of body vibrations induced by the heart beat. SCG contains information on cardiac mechanics, in particular heart sounds and cardiac output. In this paper we present a new wearable device for SCG recordings during long term monitorings, and the results of a validation test in 4 subjects. The system is based on the integration of the MagIC smart shirt (i.e., a textile-based wearable system for the assessment of ECG and respiratory movements), and an external triaxial MEMS accelerometer positioned on the left clavicle. SCG was estimated as the average of accelerations occurred in each heart beat. The SCG components due to the valve closure and to recoil forces following the heart contraction (ballistocardiogram) were extracted by high-pass (>18 Hz) and band-pass (0.6-20 Hz) filters respectively. Then the difference between the I and J waves of the ballistocardiogram ( lI-Jl index, possibly related to the cardiac output) was identified by an ad-hoc procedure and compared with the model flow indirect estimation of cardiac output. Validation on 4 volunteers showed that: 1) our wearable system provides statistically consistent estimates of both heart- sound related vibrations and recoil movements; 2) reliable estimates of the II-Jl index can be obtained by considering about 1 minute of SCG recording in stationary conditions; and 3) changes of the II-Jl index during exercise correlate well with changes of cardiac output estimated by the model flow.
  • Keywords
    accelerometers; biomechanics; biomedical electronics; cardiology; medical signal processing; microsensors; patient monitoring; MagIC smart shirt; ballistocardiogram I waves; ballistocardiogram J waves; band pass filters; external triaxial MEMS accelerometer; heart beat induced body vibrations; heart contraction recoil; high pass filters; long term heart monitoring; seismocardiogram; valve closure; wearable SCG device; Acceleration; Accelerometers; Band pass filters; Biomedical monitoring; Electrocardiography; Heart beat; Heart valves; Micromechanical devices; Testing; Vibrations; Ballistocardiography; Cardiac Output; Electrocardiography; Exercise; Humans; Kinetocardiography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE
  • Conference_Location
    Lyon
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-0787-3
  • Type

    conf

  • DOI
    10.1109/IEMBS.2007.4353199
  • Filename
    4353199