Sensitivity analysis of Heuer´s method for motion-artifacts reduction in capacitive biopotential measurements
Author
Yang, K.T. ; Serteyn, A. ; Vullings, R. ; Bergmans, J.W.M.
Author_Institution
Dept. of Electr. Eng., Eindhoven Univ. of Technol., Eindhoven, Netherlands
fYear
2014
fDate
22-24 Oct. 2014
Firstpage
192
Lastpage
195
Abstract
Capacitive biopotential measurements are susceptible to motion-artifacts (MA) due to the non-contact property of capacitive electrodes. Among the methods used to reduce MA, the two proposed by Serteyn et al. and Heuer et al. are based on analyzing the time domain behavior of the measurement system. Serteyn´s method only partially reduces MA, while Heuer´s method can completely remove MA with a reconstruction equation. However, it has not been investigated to what extent Heuer´s method can tolerate the inaccuracies in the parameters of the reconstruction equation. This is why we performed a sensitivity analysis on Heuer´s method using dynamic simulations in MATLAB with different inaccuracies added to the known reconstruction parameters. The results were compared to those of Serteyn´s method, showing that Heuer´s method becomes inferior to Serteyn´s method at a lower inaccuracy of parameters when the body static charge is higher.
Keywords
bioelectric potentials; biomedical electrodes; sensitivity analysis; Heuer method; Serteyn method; body static charge; capacitive biopotential measurements; capacitive electrodes; motion-artifact reduction; reconstruction equation; sensitivity analysis; time domain behavior; Additives; Capacitance; Electrocardiography; Electrodes; Equations; Mathematical model; Skin; Capacitive measurement; capacitive electrode; motion-artifacts;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Circuits and Systems Conference (BioCAS), 2014 IEEE
Conference_Location
Lausanne
Type
conf
DOI
10.1109/BioCAS.2014.6981695
Filename
6981695
Link To Document