• DocumentCode
    184714
  • Title

    An implementation of a wavelet-based seizure detection filter suitable for realtime closed-loop epileptic seizure suppression

  • Author

    van Dongen, M.N. ; Karapatis, A. ; Kros, L. ; Eelkman Rooda, O.H.J. ; Seepers, Robert M. ; Strydis, Christos ; De Zeeuw, Chris I. ; Hoebeek, F.E. ; Serdijn, Wouter A.

  • Author_Institution
    Sect. Bio-Electron., Delft Univ. of Technol., Delft, Netherlands
  • fYear
    2014
  • fDate
    22-24 Oct. 2014
  • Firstpage
    504
  • Lastpage
    507
  • Abstract
    This paper presents the design and implementation of a real-time epilepsy detection filter that is suitable for closed-loop seizure suppression. The design aims to minimize the detection delay, while a reasonable average detection rate is obtained. The filter is based on a complex Morlet wavelet and uses an adaptive thresholding strategy for the seizure discrimination. This relatively simple configuration allows the algorithm to run on a cheap and readily available microprocessor prototyping platform. The performance of the filter is verified using both in vivo real-time measurements as well as simulations over a pre-recorded EEG dataset (29.75 hours with 1914 seizures). An average detection delay of 492 ms is achieved with a sensitivity of 96.03% and a specificity of 93.60%.
  • Keywords
    adaptive systems; closed loop systems; delays; electroencephalography; filters; medical disorders; medical signal detection; medical signal processing; microcomputers; minimisation; prototypes; real-time systems; wavelet transforms; EEG dataset prerecording; adaptive thresholding strategy; average detection delay; average detection rate; complex Morlet wavelet; detection delay minimization; detection sensitivity; detection specificity; in vivo real-time measurement; microprocessor prototyping platform; real-time closed-loop epileptic seizure suppression; real-time epilepsy detection filter design; real-time epilepsy detection filter implementation; seizure discrimination; simulation; time 29.75 hour; time 492 ms; wavelet-based seizure detection filter; Accuracy; Delays; Electroencephalography; Epilepsy; Finite impulse response filters; In vivo; Real-time systems;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference (BioCAS), 2014 IEEE
  • Conference_Location
    Lausanne
  • Type

    conf

  • DOI
    10.1109/BioCAS.2014.6981773
  • Filename
    6981773