• DocumentCode
    1854678
  • Title

    Hydraulic analyzer for catheter-transducer systems

  • Author

    Martinez, R.J. ; Felice, Carmelo Jose ; Valentinuzzi, Max E.

  • Author_Institution
    Lab. de Bioingenieria, Univ. Nacional de Tucuman, Argentina
  • fYear
    1994
  • fDate
    3-6 Nov 1994
  • Abstract
    It is well known that during a direct blood pressure measurement when one adds a small air bubble in a catheter-transducer system (CTS) the damping coefficient increases and the natural frequency decreases. This effect deforms the pressure waveforms. Here, the authors describe an instrument that analyses the CTS by applying a sinusoidal sweep frequency and step waveforms. Furthermore, it produces physiological and pathological waveforms for control or educational purposes
  • Keywords
    biomedical equipment; blood pressure measurement; damping; catheter-transducer systems; control purpose; damping coefficient; direct blood pressure measurement; educational purpose; hydraulic analyzer; natural frequency; pathological waveforms; physiological waveforms; pressure waveforms deformation; sinusoidal sweep frequency; small air bubble; step waveforms; Blood; Blood pressure; Calibration; Catheters; Computer displays; Damping; Frequency; Frequency measurement; Instruments; Keyboards; Logic; Pathology; Performance evaluation; Pressure measurement; Testing; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE
  • Conference_Location
    Baltimore, MD
  • Print_ISBN
    0-7803-2050-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1994.412070
  • Filename
    412070