• DocumentCode
    1865031
  • Title

    Finger-force measurement-device for hand rehabilitation

  • Author

    Olandersson, Sofia ; Lundqvist, Helene ; Bengtsson, Martin ; Lundahl, Magnus ; Baerveldt, Albert-Jan ; Hilliges, Marita

  • Author_Institution
    Sch. of Bus. & Eng., Halmstad Univ., Sweden
  • fYear
    2005
  • fDate
    28 June-1 July 2005
  • Firstpage
    135
  • Lastpage
    138
  • Abstract
    The purpose was to develop an extension fingerforce measurement device, and investigate the intraindividual repeatability. The design of the measuring device allows single finger force and whole hand measurements, and the repeatability error on extension finger forces was measured, both on the whole hand, as well as on individual fingers. The tests showed that a repeatability error of less then 15 % can be achieved for single finger measurements and less then 21 % for whole hand measurements.
  • Keywords
    biomechanics; force measurement; patient rehabilitation; extension finger forces; finger-force measurement-device; hand rehabilitation; intraindividual repeatability; Arthritis; Diseases; Fingers; Force measurement; Gaussian distribution; Injuries; Muscles; Pain; Testing; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Rehabilitation Robotics, 2005. ICORR 2005. 9th International Conference on
  • Print_ISBN
    0-7803-9003-2
  • Type

    conf

  • DOI
    10.1109/ICORR.2005.1501069
  • Filename
    1501069