DocumentCode
1888985
Title
An Assessment on the Accuracy of Time-Domain Reflectometry for Measuring Level and Permittivity of Liquids
Author
Cataldo, A. ; Tarricone, L. ; Vallone, M. ; Attivissimo, F. ; Trotta, A.
Author_Institution
Dipt. Ingegneria dell´´Innovazione, Universita di Lecce
fYear
2006
fDate
24-27 April 2006
Firstpage
2332
Lastpage
2337
Abstract
In a previous paper, the authors showed that time domain reflectometry (TDR) can be successfully used for an enhanced real-time monitoring of liquids inside tanks: in one shot, determination of levels, dielectric properties and multiple interfaces in layered media are possible, thus opening challenging perspectives for several monitoring applications, especially for industrial and environmental control. For such purposes, a detailed analysis of the accuracy of the proposed measurement method is mandatory. In this paper, a metrological characterization of the method is carried out. Results are proposed, demonstrating the robustness and liability of the TDR technique
Keywords
level measurement; microwave reflectometry; permittivity measurement; time-domain reflectometry; dielectric properties; environmental control; industrial control; level measurement; microwave reflectometry; permittivity measurement; real-time liquid monitoring; reflection coefficient; time domain reflectometry; Coaxial components; Dielectric constant; Dielectric measurements; Impedance; Magnetic liquids; Permittivity measurement; Probes; Reflection; Reflectometry; Time domain analysis; Dielectric Measurements; Level Measurement; Microwave Reflectometry; Reflection Coefficient; Time Domain Reflectometry (TDR);
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 2006. IMTC 2006. Proceedings of the IEEE
Conference_Location
Sorrento
ISSN
1091-5281
Print_ISBN
0-7803-9359-7
Electronic_ISBN
1091-5281
Type
conf
DOI
10.1109/IMTC.2006.328615
Filename
4124777
Link To Document