• DocumentCode
    1902012
  • Title

    Piezoresistive geometry for maximizing microcantilever array sensitivity

  • Author

    Fletcher, Patrick C. ; Xu, Y. ; Gopinath, P. ; Williams, J. ; Alphenaar, B.W. ; Bradshaw, R.D. ; Keynton, R.S.

  • Author_Institution
    Depts. of Mech. Eng., Univ. of Louisville, Louisville, KY
  • fYear
    2008
  • fDate
    26-29 Oct. 2008
  • Firstpage
    1580
  • Lastpage
    1583
  • Abstract
    Our group has focused on the development of piezoresistive cantilevers for gas detection using a nonreaction-based method - viscous damping. This work represents a continuation of this effort by reporting on the development, fabrication and testing of a second generation device consisting of asymmetric piezoresistive cantilevers to maximize device sensitivity. The first mode of resonance, static & dynamic sensitivities, and Q-factors ranged from 40 kHz - 63 kHz, 1-6 Omega/mum & 3-17 Omega/mum, and 1700 - 4200, respectively, depending on piezoresistor geometry and beam length. The redesigned piezoresistive microcantilevers were 15 - 200 times more sensitive than the best symmetric design previously reported by our group.
  • Keywords
    cantilevers; geometry; micromechanical devices; piezoresistive devices; asymmetric piezoresistive cantilevers; device sensitivity maximization; gas detection; microcantilever array sensitivity; nonreaction-based method; piezoresistive geometry; viscous damping; Chemicals; Computational geometry; Damping; Electrical resistance measurement; Immune system; Leg; Optical resonators; Piezoresistance; Stress measurement; Testing; Piezoresistor; array; cantilever; microcantilever; optimization;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2008 IEEE
  • Conference_Location
    Lecce
  • ISSN
    1930-0395
  • Print_ISBN
    978-1-4244-2580-8
  • Electronic_ISBN
    1930-0395
  • Type

    conf

  • DOI
    10.1109/ICSENS.2008.4716751
  • Filename
    4716751