    Radiated immunity measurement comparison between OSL and chamber

  • Author

    Jang, Tae Heon ; Cho, Won Seo

  • Author_Institution
    EMC Team, Korea Testing Lab., Seoul
  • fYear
    2006
  • fDate
    Feb. 27 2006-March 3 2006
  • Firstpage
    477
  • Lastpage
    480
  • Abstract
    In CISPR 20, radiated immunity test is performed in the open stripline (OSL) which height is 0.8 m. CISPR 20 also specifies that equipment not fitting inside open stripline shall be measured according to IEC 61000-4-3 in frequency range of 80 MHz to 150 MHz with the same limits as in the OSL. In this paper, the analysis of the field distribution in the OSL was completed by moment method and the testing according to both methods was carried out and the immunity level has compared. The result shows that the testing in chamber is more severe than in OSL
  • Keywords
    anechoic chambers (electromagnetic); immunity testing; method of moments; strip lines; 80 to 150 MHz; OSL; moment method; open stripline; radiated immunity measurement; radiated immunity test; Conductors; Electromagnetic analysis; Electromagnetic compatibility; Frequency; IEC standards; Immunity testing; Laboratories; Moment methods; Stripline; TV receivers;