• DocumentCode
    191278
  • Title

    Space distribution of SAR and temperature in human body model with tumor using waveguide applicator array

  • Author

    Ondrej, Fiser ; Ilja, Merunka ; Lucie, Vojackova ; Jan, Vrba

  • Author_Institution
    Dept. of Electromagn. Field, Czech Tech. Univ. in Prague, Prague, Czech Republic
  • fYear
    2014
  • fDate
    6-9 Oct. 2014
  • Firstpage
    484
  • Lastpage
    487
  • Abstract
    The way to treat tumors in head and neck region is the main aim of this work. For this purpose we used array of waveguide applicators working on frequency 434 MHz. Characteristics of used applicator array tested through the EM field simulator SEMCAD X on numerical model and model made from CT scans. We observed SAR and temperature distribution in tumor and sensitive tissues. For better results we optimized SAR distribution and tried to focus energy to tumor. This contribution shows good control of SAR distribution patterns and offers a solution for patients with large tumors.
  • Keywords
    computerised tomography; hyperthermia; numerical analysis; patient treatment; temperature distribution; tumours; CT scans; EM field simulator SEMCAD X; SAR distribution; head region; human body model; neck region; numerical model; sensitive tissues; space distribution; temperature distribution; tumor treatment; waveguide applicator array; Applicators; Arrays; Glands; Hyperthermia; Neck; Temperature distribution; Tumors; Antenna Array; Head and Neck Treatment; Hyperthermia; SAR Optimalization; Thermal Simulation; Treatment Planning; Waveguide;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference (EuMC), 2014 44th European
  • Conference_Location
    Rome
  • Type

    conf

  • DOI
    10.1109/EuMC.2014.6986476
  • Filename
    6986476