• DocumentCode
    1995727
  • Title

    Fabrication of 20 MHz convex array transducers for high frequency ophthalmic imaging

  • Author

    Kim, Hyung Ham ; Cannata, Jonathan M. ; Williams, Jay A. ; Chang, Jin Ho ; Shung, K. Kirk

  • Author_Institution
    Dept. of Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA
  • fYear
    2009
  • fDate
    20-23 Sept. 2009
  • Firstpage
    1130
  • Lastpage
    1133
  • Abstract
    20 MHz 192-element convex array transducers have been designed, fabricated and tested. Major design parameters are 111 ¿m pitch, 24 mm radius of curvature, 52° maximum view angle, 7 mm elevation width, and 30 mm geometric focus. A 1-3 composite was fabricated with a ceramic post width of 28 ¿m and a polymer kerf width of 9 ¿m. The ceramic volume fraction was approximately 57%. Three ceramic posts formed one element in the azimuth direction. A custom designed flexible circuit was aligned and attached to a scratch-diced composite and 75 ¿ micro-coaxial cables. A single epoxy matching layer and a backing block were used for acoustic matching. The pulse-echo response for a typical element acquired at the axial peak, 28.0 mm, matched with a designed geometrical focusing at 30 mm, had a center frequency of 24.4 MHz and 64% of -6 dB fractional bandwidth. The estimated insertion loss at 24 MHz after compensating for attenuation and reflection was 35 dB and the maximum crosstalk was -17 dB and -22 dB for the nearest and for the next-nearest element respectively.
  • Keywords
    biomedical transducers; biomedical ultrasonics; ceramics; coaxial cables; composite materials; eye; flexible electronics; interdigital transducers; polymers; ultrasonic transducer arrays; acoustic matching; azimuth direction; backing block; ceramic posts; ceramic volume fraction; convex array transducers; designed flexible circuit; frequency 20 MHz; frequency 24 MHz; frequency 24.4 MHz; high frequency ophthalmic imaging; maximum crosstalk; microcoaxial cables; polymer kerf width; pulse-echo response; resistance 75 ohm; scratch-diced composite; single epoxy matching layer; size 24 mm; size 24 mum; size 28 mum; size 7 mm; size 9 mum; Azimuth; Cables; Ceramics; Fabrication; Flexible printed circuits; Focusing; Frequency; Polymers; Testing; Transducers; 1–3 composite; convex array; curved linear array; high frequency ultrasound; ophthalmic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2009 IEEE International
  • Conference_Location
    Rome
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4244-4389-5
  • Electronic_ISBN
    1948-5719
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2009.5441610
  • Filename