    Verification of an image calibration method in ultrasound contrast agent imaging on a perfusion phantom

  • Author

    Finfer, Daniel C. ; Tang, Meng-Xing ; Eckersley, Robert J.

  • Author_Institution
    Dept. of Bioeng., Imperial Coll. London, London, UK
  • fYear
    2009
  • fDate
    20-23 Sept. 2009
  • Firstpage
    267
  • Lastpage
    270
  • Abstract
    An image calibration method, based on ¿pulse inversion ratio¿ (PIR) was tested using a flow phantom consisting of a dialysis cartridge to simulate tissue perfusion. By comparing PIR to B-mode and PI output, it was seen that PIR is sensitive to the flow of contrast agents, and is capable of suppressing shadowing artifacts.
  • Keywords
    biological tissues; biomedical ultrasonics; calibration; haemorheology; patient diagnosis; phantoms; dialysis cartridge; image calibration method; perfusion phantom; pulse inversion ratio; shadowing artifact suppression; tissue perfusion; ultrasound contrast agent imaging; Attenuation; Biomedical engineering; Calibration; Circuits; Educational institutions; Imaging phantoms; Optical fiber testing; Pulse inverters; Silicon; Ultrasonic imaging; calibration; contrast agents; perfusion;