• DocumentCode
    2052417
  • Title

    Performance results of a prototype depth-encoding PET detector

  • Author

    Meng, L.J. ; Ramsden, D.

  • Author_Institution
    Dept. of Phys. & Astron., Southampton Univ., UK
  • Volume
    2
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    929
  • Abstract
    The results of a feasibility study for a high-resolution PET detector system are presented. The detector is based on the use of etched LSO scintillation crystals viewed from just one end. The authors have demonstrated that this configuration yields Depth-of-Interaction information whilst also preserving good energy and coincidence-time resolution. Prototype detectors have been built using 6×6 crystal arrays coupled to 61-pixel hybrid photodiodes. The measurements demonstrate a DOI resolution of 5~7 mm in a 30 mm long detector. The signal level was measured to be in the range 250 to 600 photoelectrons depending on the location of the 511 keV energy deposit in the crystal. A resolving-time of less than 5 ns was also achieved. The technique depends on the assumption that all of the events detected above a selected threshold value are, in fact, full-energy deposits. The validity of this assumption and the contribution from photons scattered in the object, have been carefully studied using a well-calibrated Monte Carlo simulation of the proposed small animal PET imager
  • Keywords
    Monte Carlo methods; biomedical equipment; gamma-ray detection; image coding; image resolution; photodiodes; positron emission tomography; 30 mm; 5 ns; 511 keV; 61-pixel hybrid photodiodes; medical diagnostic imaging; medical instrumentation; nuclear medicine; photoelectrons; prototype depth-encoding PET detector; resolving-time; small animal PET imager; well-calibrated Monte Carlo simulation; Crystals; Energy resolution; Etching; Photodiodes; Positron emission tomography; Prototypes; Sensor arrays; Signal detection; Signal resolution; Solid scintillation detectors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.845815
  • Filename
    845815