DocumentCode
2077038
Title
Design, fabrication, and characterization of an electrochemically-based dose tracking system for closed-loop drug delivery
Author
Sheybani, Roya ; Cabrera-Munoz, N.E. ; Sanchez, Tonametl ; Meng, Ellis
Author_Institution
Dept. of Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA
fYear
2012
fDate
Aug. 28 2012-Sept. 1 2012
Firstpage
519
Lastpage
522
Abstract
A real-time integrated electrochemically-based dose tracking system for closed-loop drug delivery is presented. Thin film Pt sensors were integrated in an electrolytic MEMS drug delivery pump to allow dose tracking via electrochemical impedance measurement. Measurement electrode placement and composition were investigated. A bolus resolution of 230 nL was demonstrated. The sensor was calibrated for use with water (low conductivity) and 1 × PBS (high conductivity), the selected model aqueous drugs. The impedance response is dependent on delivered volume and not affected by actuation parameters. A graphical user interface was created for real-time impedance based dose tracking and leakage/blockage detection in the system. Drift in the impedance response of an idle system after perturbation (actuation) were investigated and mitigated through the use of Pt wire electrodes as opposed to thin film electrodes.
Keywords
bioMEMS; biomedical electrodes; biomedical measurement; calibration; dosimetry; drug delivery systems; electrochemical electrodes; electrochemical impedance spectroscopy; electrolytic devices; graphical user interfaces; medical computing; microactuators; microfabrication; micropumps; microsensors; platinum; thin film sensors; Pt; actuation parameters; bolus resolution; calibration; closed-loop drug delivery; electrochemical impedance measurement; electrolytic MEMS drug delivery pump; graphical user interface; impedance response; leakage-blockage detection; measurement electrode composition; measurement electrode placement; model aqueous drugs; real-time impedance based dose tracking; real-time integrated electrochemically-based dose tracking system; thin film electrodes; thin film sensors; wire electrodes; Bellows; Drug delivery; Drugs; Electrochemical processes; Electrodes; Fluids; Impedance; Computer Systems; Drug Delivery Systems; Electric Impedance; Electrochemical Techniques; Electrodes; Electrolysis; Equipment Design; Humans;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Conference_Location
San Diego, CA
ISSN
1557-170X
Print_ISBN
978-1-4244-4119-8
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/EMBC.2012.6345982
Filename
6345982
Link To Document