DocumentCode
2091885
Title
Short and long term stability of HP CS standards including environmental effects
Author
Kalliomaki, K. ; Mansten, T. ; Mannermaa, J.
Author_Institution
Electr. Group, MIKES, Espoo
fYear
2004
fDate
5-7 April 2004
Firstpage
409
Lastpage
410
Abstract
The Finnish Centre for Metrology and Accreditation (MIKES) has two HP 5071A Cs-beam standards (standard version from 1997 and high stability version from 2000) and one CH1-76 Passive hydrogen maser (since February 2003). A special four-channel phase comparator with better than 0.3 ps single shot resolution is used to study the short-term stability of the Cs-standards, with hydrogen maser as a reference. Long term stability is measured with GPS as a reference. Allan stability plots indicate that the high stability version 5071A is approximately three times better than the standard version. The high stability version has only slightly smaller long term drift (ageing) than standard version but frequency retrace after clock failure or power blackout is significantly better in high stability version.
Keywords
atomic clocks; masers; measurement standards; phase comparators; stability; Allan stability; CH1-76 Passive hydrogen maser; Finnish Centre for Metrology and Accreditation; GPS; HP 5071A caesium-beam standards; Long term stability; clock failure; environmental effects; four-channel phase comparator; frequency retrace; power blackout; short-term stability; Allan stability; Cesium clock; Environmental effects;
fLanguage
English
Publisher
iet
Conference_Titel
Frequency and Time Forum, 2004. EFTF 2004. 18th European
Conference_Location
Guildford
ISSN
0537-9989
Print_ISBN
0-86341-384-6
Type
conf
Filename
5074984
Link To Document