• DocumentCode
    2094685
  • Title

    Multi-lead QRS detection using window pairs

  • Author

    Torbey, S. ; Akl, Selim G. ; Redfearn, D.P.

  • Author_Institution
    Queen´s Univ., Kingston, ON, Canada
  • fYear
    2012
  • fDate
    Aug. 28 2012-Sept. 1 2012
  • Firstpage
    3143
  • Lastpage
    3146
  • Abstract
    We designed a novel approach for multi-lead QRS detection. The algorithm uses one equation with two different window widths to generate a feature signal and a detection threshold. This enables it to adapt to various changes in QRS morphology and noise levels, resulting in a detection error rate of just 0.29% on the MIT-BIH Arrhythmia Database. The algorithm is also computationally efficient and capable of resolving differences between multiple leads by automatically attaching a confidence value to each QRS detection.
  • Keywords
    electrocardiography; medical signal detection; medical signal processing; MIT-BIH Arrhythmia Database; QRS morphology; confidence value; detection threshold; feature signal generation; multilead QRS detection; noise level; window pairs; window width; Databases; Detection algorithms; Electrocardiography; Feature extraction; Morphology; Noise; Rhythm; Algorithms; Arrhythmias, Cardiac; Electrocardiography; Humans; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4119-8
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2012.6346631
  • Filename
    6346631