DocumentCode
2100071
Title
Enhancing the sensitivity of miniaturized quadrupole mass spectrometers
Author
Wilamowski, Bodgan ; Bell, David ; Legowski, Stanislaw ; Abdelkader, Yaser
Author_Institution
Graduate Center, Univ. of Idaho, Boise, ID, USA
Volume
1
fYear
2001
fDate
2001
Firstpage
147
Abstract
A novel two stage mass filter where the first stage operates in DC/RF mode and second in RF only mode is proposed. This design can be characterized by very high mass resolution and may not require high precision mechanical and electronic design. An ion detector by integrating a VLSI chip with a Faraday cup is described. Extremely low biasing currents, differential and narrow frequency band operation resulted in significant improvement of the signal to noise ratio
Keywords
Faraday effect; VLSI; counters; mass spectrometer accessories; mass spectrometers; DC/RF mode operation; Faraday cup; RF mode operation; VLSI chip; differential frequency band operation; high mass resolution; ion detector; low biasing currents; narrow frequency band operation; sensitivity enhancement; signal to noise ratio improvement; two stage mass filter; Detectors; Educational institutions; Equations; Filters; Mass spectroscopy; Radio frequency; Signal resolution; Signal to noise ratio; Very large scale integration; Voltage;