  • Abstract
    An impedance bridge that compares two-terminal standard inductors to characterized resistors is described. A dual-channel digitally synthesized source and sampling digital multimeter are used to generate and measure relevant bridge signals. A linear interpolation algorithm is used to autocalibrate the bridge to a 1 nF air-dielectric capacitor. An intercomparison of the new bridge with existing measurement standards in the low audio frequency range for inductors 1 mH to 10 H is reported.<>
  • Keywords
    automatic test equipment; bridge instruments; calibration; digital multimeters; electric impedance measurement; inductors; interpolation; measurement standards; NIST digital impedance bridge; air-dielectric capacitor; audio frequency range; autocalibration; bridge signals; dual-channel digitally synthesized source; intercomparison; linear interpolation algorithm; measurement standards; sampling digital multimeter; two-terminal standard; Bridges; Capacitors; Impedance; Inductors; Interpolation; NIST; Resistors; Sampling methods; Signal generators; Signal synthesis;