    Accuracy comparison of micro-angiographic detector and image intensifier for an interventional localization task

  • Author

    Zhou Wang ; Rudin, Stephen ; Wu, Ye ; Hoffmann, Kenneth R. ; Bednarek, Daniel R.

  • Author_Institution
    Dept. of Physiol. & Biophys., State Univ. of New York, Buffalo, NY, USA
  • fYear
    2002
  • fDate
    2002
  • Firstpage
    553
  • Lastpage
    556
  • Abstract
    In order to assess the effect of the use of a high-resolution micro-angiographic detector on the accuracy of performing interventional localization tasks, we set up a simple task to localize a guide wire tip relative to one edge of a thin marker wire. Trained observers were asked to evaluate the micro-angiographic images as well as the comparison standard x-ray image intensifier (XRII) images, both of which were taken with the guidewire in the same positions. The observers were able to get more accurate estimation results using images obtained with the micro-angiographic detector than the XRII; the average distance error and performance variability both were reduced by factors of 2-3.
  • Keywords
    CCD image sensors; X-ray apparatus; biomedical equipment; blood vessels; diagnostic radiography; image intensifiers; image resolution; patient treatment; CCD camera; accuracy comparison; average distance error; comparison standard x-ray image intensifier images; guide wire tip; guidewire; high-resolution micro-angiographic detector; image intensifier; interventional localization task; performance variability; thin marker wire; Aneurysm; Biophysics; Detectors; High-resolution imaging; Image edge detection; Image intensifiers; Physiology; Spatial resolution; Wire; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Imaging, 2002. Proceedings. 2002 IEEE International Symposium on
  • Print_ISBN
    0-7803-7584-X
  • Type

    conf

  • DOI
    10.1109/ISBI.2002.1029317
  • Filename
    1029317