Abstract
The optically pumped Cs beam frequency standard under development in our laboratory was first evaluated in 1993. The short term stability was /spl sigma//sub y/(/spl tau/) is 5.5/spl middot/10/sup -13//spl tau//sup - 1/2 /. The accuracy has been estimated to 1.1/spl middot/10/sup -13/, mainly limited by the distributed phase shift. After modification of the cavity, a new evaluation is in progress and is presented.<>
Keywords
Doppler effect; Zeeman effect; atomic beams; caesium; frequency measurement; measurement standards; optical pumping; Cs; Doppler shift; LPTF; Ramsey cavity; Zeeman effect; cavity pulling; optically pumped Cs beam frequency standard; primary frequency standard; Frequency; Laboratories; Laser excitation; Optical beams; Optical pumping; Optical variables control; Phase estimation; Stability; Structural beams; Uncertainty;