• DocumentCode
    2187795
  • Title

    Error corrected impedance measurements with a network analyzer

  • Author

    Heuermann, H. ; Schiek, B.

  • Author_Institution
    Inst. fur Hochfrequenztech., Ruhr-Univ., Bochum, Germany
  • fYear
    1994
  • fDate
    June 27 1994-July 1 1994
  • Firstpage
    125
  • Lastpage
    126
  • Abstract
    For a network analyzer some error corrected impedance or two-port measurements are presented. These ZU and YU self-calibration procedures only need one known and one or two unknown impedances for the calibration process. The good performance of these simple methods is demonstrated by experimental results.<>
  • Keywords
    calibration; electric impedance measurement; measurement errors; microwave measurement; multiport networks; network analysers; calibration; error corrected impedance measurement; network analyzer; self-calibration; two-port measurements; Calibration; Equations; Error analysis; Error correction; Impedance measurement; Microwave measurements; Performance analysis; Performance evaluation; Switches; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 1994. Digest., 1994 Conference on
  • Conference_Location
    Boulder, CO, USA
  • Print_ISBN
    0-7803-1984-2
  • Type

    conf

  • DOI
    10.1109/CPEM.1994.333421
  • Filename
    333421